2
EXP. COMPARISON FUNCTIONS
118
EXP
. COMPARISON
FUNCTIONS
Chapter
2
* The Hysteresis setting is not reflected in the Difference-HI/LO Limit comparison operation.
* If the Comparison Timing condition is not met, the comparison result of the value targeted for the
Comparison Timing turns OFF.
* OK is displayed each on the LCD screen as hold value A is OK, and hold B and the difference
value are OK.
2-3.
Relative Value Comparison
(in expansion double hold only)
ON / OFF
Initial value: OFF
How to set
On two hold points (A and B) in Double hold, HI/LO Limit comparisons of the comparison
standard hold value, and HI/LO Limit judgments of the difference (relative value) between the
comparison standard hold value and the other hold value can be made.
●
In the case of
Comparison Standard A
where
Hold A: Sample
Hold B: Inflection Point
* The comparison output
in the timing chart
shown on the right-hand
side applies to the case
where the Comparison
Timing setting in
Comparison Setting is
All Time.
FCN-365P024-AU/FCN-360C024-B
External I/O
A1
-
COM1
COM2
-
B1
A2
OUT
HI-B & DF-HI
WORK0
IN
B2
A3
OUT
HI-A
WORK1
IN
B3
A4
OUT
OK
WORK2
IN
B4
A5
OUT
LO-A
WORK3(LOCK)
IN
B5
A6
OUT
LO-B & DF-LO
T/H
IN
B6
A7
-
COM1
COM2
-
B7
A8
OUT
HOLD END(H/E)
SECTION
IN
B8
A9
OUT
RUN
D/Z
IN
B9
A10
OUT
EVENT
GRAPH TRIG
IN
B10
A11
OUT
SIF
CAL0
IN
B11
A12
OUT
SIF
CAL1
IN
B12
- External I/O pin assignments for Comparison Standard B
SET.
Exp. Comp. Set.
Page 1
→
→
Exp. Setting
→
SECTION
T/H
ON
H/E
ON
HI-A
ON
OK
ON
LO-A
ON
RL-HI
ON
RL-LO
OFF
HI-A
LO-A
Hold Section
Relative-HI
Relative-LO
A comparison of Hold B
is made based on the
sample hold point of Hold A.
Relative hold value B
from Hold A
Summary of Contents for DS-6200
Page 1: ......
Page 11: ...CONTENTS 10 M E M O...
Page 12: ...11 11 STANDARD...
Page 13: ...12 12 M E M O...
Page 92: ...91 91 EXPANSION...
Page 93: ...92 92 M E M O...
Page 95: ...Expansion Function 94 M E M O...