1
EXP. HOLD FUNCTIONS
113
EXP. HOLD
FUNCTIONS
Chapter
1
1-6.
Hold End Timing
The HOLD END (H/E) signal timing when Inflection Point, Relative Maximum, Relative
Minimum or Relative Difference hold is used can be changed.
Hold Sct / Detect Sct
Initial value: Detect Sct
How to set
- Hold Sct:
HOLD END (H/E) is on according to timing in which the SECTION signal is
off.
- Detect Sct:
HOLD END (H/E) is on immediately after the detection of the hold point.
■
Hold operations
SET.
Exp. Hold Set.
Page 2
→
→
Exp. Setting
→
In the case of Inflection Point hold
SECTION
T/H
Tracking Section
Detection
Hold Section
When the HOLD END timing is "Hold section"
H/E
HOLD Status
OFF
ON
OFF
INF Status
OFF
Blink
ON
OFF
When the HOLD END timing is "Detection section"
H/E
HOLD Status
OFF
ON
OFF
INF Status
OFF
Blink
ON
OFF
Tracking Section
Indicated value
Sensor Input Value
Summary of Contents for DS-6200