5 Function Setting
LF(R) Series Operation Manual
- 70 -
5-2 Readability Setting
Use this function to set the readability. The larger the readability becomes, the less the
balance is affected by external influences. In addition, it takes less time for the balance
reading to stabilize.
Each unit has a different readability.
The readability setting is available for the single-range type only.
1
Go to the function setting mode.
(Refer to Section 2-6 “Basics of Function
Settings.”)
Press and hold the [Function] key until “Func” is
displayed.
2
Select the readability.
Press the [Function] key several times to select “b2.d.A.”
Press the [Zero/Tare] key to select from 1
through 5.
3
Save the setting.
Press the [Set] key.
The balance returns to the weight display.
To set the readability of unit b, select “b4.d.b” in step 2.
You can also set the same unit for unit A and unit b, and set different minimum indications,
so that unit A and unit b can be used to switch the readability.
List of readability settings
Set value
LF(R)-124 / -224
LF-135R
Unit mg
g
ct mom
mg
g
ct
mom
1 0.1
0.0001
0.001
0.0001
0.01
0.00001
0.0001
0.00001
2 0.2
0.0002
0.002
0.0002
0.2
0.00002
0.0002
0.00002
3 0.5
0.0005
0.005
0.0005
0.05
0.00005
0.0005
0.00005
4 1
0.001
0.01
0.001
0.1
0.0001
0.001
0.0001
5 2
0.002
0.02
0.002
0.2
0.0002
0.002
0.0002
Summary of Contents for LF Series
Page 11: ... x Memo ...
Page 131: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 132: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 133: ...Appendixes LF R Series Operation Manual 122 ...
Page 134: ...LF R Series Operation Manual Appendixes 123 ...
Page 135: ...Appendixes LF R Series Operation Manual 124 ...
Page 136: ...LF R Series Operation Manual Appendixes 125 ...
Page 137: ...Appendixes LF R Series Operation Manual 126 ...
Page 138: ...LF R Series Operation Manual Appendixes 127 ...
Page 139: ...Appendixes LF R Series Operation Manual 128 ...
Page 140: ...LF R Series Operation Manual Appendixes 129 ...
Page 145: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 146: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 150: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 151: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 154: ...LF R Series Operation Manual Index for Terms 143 Memo き ...