6 Input/Output to/from External Devices
LF(R) Series Operation Manual
- 102 -
Meanings of data in special format 2
[S1 to S3] (three characters)
Indicates stable/unstable.
S1 S2 S3
S1
code
S2
code
S3
code
Meaning
S (SP)
S 53H 20H 53H Stable
S (SP)
D 53H 20H 44H Unstable
[SP, D1 to D10]
Stores numeric data.
D1 to D10
Code
Description
0 to 9
30H to 39H
0 to 9 (numeric) right-aligned
2EH
Decimal point (floating)
−
2DH
Polarity negative “
−
”
SP (Space)
20H
Polarity positive “+” or zero
Spaces at the front and back of a numeric value
Spaces in the digit places where there is no
numeric value
[U1, U2, U3] (three characters)
Indicates the unit used to show numeric data.
This is variable-length data and the number of characters varies between 1 to 3.
Summary of Contents for LF Series
Page 11: ... x Memo ...
Page 131: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 132: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 133: ...Appendixes LF R Series Operation Manual 122 ...
Page 134: ...LF R Series Operation Manual Appendixes 123 ...
Page 135: ...Appendixes LF R Series Operation Manual 124 ...
Page 136: ...LF R Series Operation Manual Appendixes 125 ...
Page 137: ...Appendixes LF R Series Operation Manual 126 ...
Page 138: ...LF R Series Operation Manual Appendixes 127 ...
Page 139: ...Appendixes LF R Series Operation Manual 128 ...
Page 140: ...LF R Series Operation Manual Appendixes 129 ...
Page 145: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 146: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 150: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 151: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 154: ...LF R Series Operation Manual Index for Terms 143 Memo き ...