LF(R) Series Operation Manual
3 Various Measuring Methods
- 43 -
Detailed function settings
The limit function can be set to a specific value with the function settings.
When the function setting “2.SEL” is “2” or “3,” the following items can be set by pressing the
[Function] key. Set these items as required.
☆
Initial setting
Condition 21.Co.
☆
1: Always judge.
2: Judge only when the balance is stable.
Range to cover
22.Li
☆
0: Detect when the limit is exceeded by more than five
divisions.
1: Detect when the limit is exceeded by more than fifty
divisions
2: Detect both when the limit is exceeded and when it is
not reached.
Point scale
23.Pi
☆
0: 1-point scale (OK and LO are judged.)
1: 1-point scale (HI and OK are judged.)
2: 2-point scale (HI, OK, and LO are judged.)
3: 3-point scale (Ranks 1 to 4 are judged.)
4: 4-point scale (Ranks 1 to 5 are judged.)
Judge using
24.tYP.
☆
1: Judge using definite values
2: Judge using deviation values
Buzzer for rank 1
25.bu.1
☆
0: The buzzer is not sounded for rank 1 (LO).
1: The buzzer is sounded for rank 1 (LO).
Buzzer for rank 2
26.bu.2
☆
0: The buzzer is not sounded for rank 2 (OK).
1: The buzzer is sounded for rank 2 (OK).
Buzzer for rank 3
27.bu.3
☆
0: The buzzer is not sounded for rank 3 (HI).
1: The buzzer is sounded for rank 3 (HI).
Buzzer for rank 4
28.bu.4
☆
0: The buzzer is not sounded for rank 4.
1: The buzzer is sounded for rank 4.
Buzzer for rank 5
29.bu.5
☆
0: The buzzer is not sounded for rank 5.
1: The buzzer is sounded for rank 5.
Separate limit values can be saved for each weighing mode. However, limit values
of both definite and deviation values cannot be saved in the same weighing mode.
Limit values can be set only in measurement mode display (Cannot be set in other
mode displays such as when a cumulative value is displayed).
Adjust the zero-point or set a tare range, as required, before setting limit values.
If the limit value entries are not lined up in the order of magnitude, three “
” will
light up. Enter the values again.
Summary of Contents for LF Series
Page 11: ... x Memo ...
Page 131: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 132: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 133: ...Appendixes LF R Series Operation Manual 122 ...
Page 134: ...LF R Series Operation Manual Appendixes 123 ...
Page 135: ...Appendixes LF R Series Operation Manual 124 ...
Page 136: ...LF R Series Operation Manual Appendixes 125 ...
Page 137: ...Appendixes LF R Series Operation Manual 126 ...
Page 138: ...LF R Series Operation Manual Appendixes 127 ...
Page 139: ...Appendixes LF R Series Operation Manual 128 ...
Page 140: ...LF R Series Operation Manual Appendixes 129 ...
Page 145: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 146: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 150: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 151: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 154: ...LF R Series Operation Manual Index for Terms 143 Memo き ...