Aoralscan 3_Manual_V1.0.0.30 - 3 -
3) After using scanner head, use alcohol to wipe and then use autoclave to sterilize it. (121°C,
102.9kPa
for 30 minutes; 134°C,
205.8kPa
for 4 minutes). Use alcohol to wipe the scanner cover.
Use dust-proof cloth to wipe the scanning window to ensure the window keeps dry.
4) Operating temperature: 10°C to 40°C, relative humidity: 30%
–
75%.
5) Storage/transport temperature: -25°C to 60°C, relative humidity: 30%
–
75%.
6) Air pressure: 86 kPa
–
106 kPa.
Note
The temperature and humidity and atmospheric pressure conditions for
storage/transportation are reflected on the outer packaging.
VII. Production date, use by date or expiration date
The production date is shown on the product label. Use period: 8 years.
Eight, the list of accessories, including accessories, accessories, wear and tear replacement
cycle and instructions on how to replace.
Scan head as a wear and tear products can be autoclaved up to 20 times, after which it needs
to be replaced. Replacement method is as follows.
(1) Disconnect the scanner power, hold the scanning head firmly with thumb and index finger
on both sides, and then gently slide the scanning head out of the scanner as shown in the figure.
(2) Hold the scan head firmly with your thumb and index finger on either side and gently
attach the scan head to the scanner with the head facing down.
CAUTION
Do not place your fingers on the lens of the scan head when removing and attaching the scan
head, because this might cause damage to the lenses.
(3) Try to gently shake the scan head to ensure that it locks into place and is stable.
CAUTION
The Aoralscan 3 intraoral scanner should not be used in close proximity or stacked with other
equipment, and if it must be used in close proximity or stacked, observe to verify proper operation
in the configuration in which it is used.
Class A equipment is intended for use in industrial environments where it may be potentially
difficult to ensure electromagnetic compatibility in other environments due to conducted and
radiated disturbances from the Aoralscan 3 intra-oral scanner.