Aoralscan 3_Manual_V1.0.0.30 - 20 -
performance of this equipment could result.
⚫
If higher IMMUNITY TEST LEVELS than those specified in IEC60601-1-2.
⚫
The minimum separation distance may be lowered. Lower minimum separation distances
shall be calculated using the equation specified in IEC60601-1-2 Chapter 8.10.
⚫
WARNING: Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation.
⚫
WARNING: Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and result in improper operation.
⚫
Note The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
Biological Safety
Meets biological criteria: ISO10993-5: 2009 (Biological evaluation of medical devices
—
Part
5:Tests for in vitro cytotoxicity); ISO10993-10: 2010 (Biological evaluation of medical devices
—
Part 10: Tests for irritation and skin sensitization).
Laser Protection
This product is a class 1 laser product and is only for maintenance, replacement and removal
by professional personnel of the manufacturer or its designated agent (if necessary). If the device
is not used, removed or replaced as required, the normal use of the device may be affected and
laser radiation may occur. If a laser component is faulty, contact the manufacturer for help.
This product is a class 1 laser product according to "IEC 60825-1:2014 Safety of laser products-
Part 1: Equipment classification and requirements", without harmful laser radiation. Users will not
be exposed to laser radiation if they operate the equipment correctly according to the instructions.
Users should be aware of optical radiation protection. Bright light is projected from the
scanning head during scanning. As with other light, there may be a temporary reduction in vision
or visual residuals. Do not look directly into the light projected by the scanning head or shine the
light into the eyes of others.