Aoralscan 3_Manual_V1.0.0.30 - 35 -
When the message disappears, the warm-up is done. The scanner is now ready for an
acquisition.
NOTES
⚫
The heater maintains constant temperature on the scanner head.
⚫
The scanner head is being heated whenever power is supplied, even if the scanner is in
standby or sleep mode.
⚫
If the heater does not reach the necessary temperature for preventing condensation
during scanning, the message of “The scanner is
pre-