Aoralscan 3_Manual_V1.0.0.30 - 30 -
(4) Connect the USB 3.0 downstream port on the pedestal and your computer with the
supplied USB 3.0 cable.
(5) Turn on the power switch on the pedestal.
(6) Click the shortcut icon of IntraoralScan on the desktop to launch the software.
Disconnecting the Scanner
CAUTION
Do not attempt to directly disconnect the scanner by removing the power cable and USB
cable.
Follow the steps below to safely disconnect the scanner:
(1) Quit the IntraoralScan scanning software.
(2) Disconnect the scanner USB 3.0 cable from the relay box.
(3) Disconnect the other port of the relay box from the computer.
(4) Right-
click the “Safely Remove Hardware” icon on Windows
taskbar and select
“
Eject Flash
Drive”.
(5) Unplug the USB flash drive and keep it in a safe place for future use.
(6) Unplug the power adapter from the wall outlet and remove the power plug from the
power connector on the pedestal.
CAUTION
Do not roll the cable over the handle of the scanner or even create any sharp bends in the
cable after you disconnect the scanner. See USB cable storage in "4.2.3 Relay box" for more details.
Calibrating the Scanner
Under these circumstances, we recommend that you shall execute the calibration for the
scanner to ensure the accuracy of scanned data:
⚫
The initial setup of the scanner is completed.
⚫
The scanner has been used for a period of time (e.g. 2 weeks).
⚫
The scanner is accidentally dropped.
⚫
Scanner brightness adjustment is recommended once every 3 months.
⚫
The calibration plate inside the calibrator is a high-precision component, and the surface
of the plate must be kept clean. If the calibration does not proceed smoothly, check the condition
of the calibration plate surface. If you find a dirty surface, consult the manufacturer or the