Aoralscan 3_Manual_V1.0.0.30 - 45 -
with the pedestal first.
⚫
Ensure that the scanner head is detached from the scanner, and the protection cap is
put on the scanner when disinfecting the scanner body.
Follow the steps below to complete the disinfection:
(1) Disconnect the power of the
Aoralscan 3
(see Disconnecting the scanner on chapter 4).
(2) Hold the scanner head firmly with your thumb and forefinger on both sides, and then
gently slide the head off from the scanner, as illustrated.
NOTE
Do not place your finger(s) on the mirror of the head when detaching as this may result in
damage to the mirror.
(3) Store the detached head in a safe place, e.g. a dental instrument tray, prior to disinfecting
the scanner body.
(4) Hold the supplied protection cap with the triangle mark facing upward. Then, align the
protection cap blocks to the matching slots on the front end of the scanner body.
(5) Slide the protection cap onto the scanner to prevent damage and dust.
CAUTIONS
⚫
When the scanner head is detached, always protect the subtle units and the inner
optical components on the front end of the scanner by putting on the supplied protection cap.
⚫
Do not attempt to clean the outer units and inner optical components on the front end
of the scanner with any sharp objects or other such tools, which may result in scratches and
damage to the scanner.
(6) Hold the scanner body with your hand.
(7) Use new cotton gauze moistened with 70%-75% solution of ethanol to wipe the surface
of scanner body.
CAUTIONS
⚫
Avoid using detergent of any kind as some detergents or surfactants might penetrate
into the surface of the scanner body.
⚫
Do not clean the intake and exhaust vents with any sharp objects or other such tools.
(8) When done, store the scanner body in a clean and safe place.