Aoralscan 3_Manual_V1.0.0.30 - 17 -
EC 61000-4-6
Radiated RF
IEC 61000-4-3
6 Vrms
ISM bands between 150
kHz and
80 MHz
80 MHz to 2.7 GHz
3V/m
6 Vrms
ISM bands between
150 kHz and 80 MHz
3 V/m
IEC 60601-1-2: 2014
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
IEC 60601-1-2: 2014
d=2.0 √P 80 MHz to 2.7 GHz
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and
d
is the recommended
separation distance in
meters (m).
Field strengths from fixed
RF
transmitters,
as
determined
by
an
electromagnetic
site
survey
a
, should be less
than the compliance level
in each frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with following symbol:
Guidance and
manufacturer’s declaration–
electromagnetic immunity
Guidance and Manufacturer's Statement - Electromagnetic emission
Aoralscan 3 is intended to be used in the following electromagnetic environment. The purchaser
or user of Aoralscan 3 should ensure that it is used in this electromagnetic environment:
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic environment
–
guidance
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.