KeyPad
Version 1.0
SCP SCIENCE
59
SPECIFICATIONS
Digi
PREP HT
ENVIRONMENTAL
Relative humidity
30% to 80%
Altitude
Up to 200m
Installation
Category ll
Pollution
Degree 2
Ambient operating
temp.
-5°C to 40°C
ELECTRICAL
Model
Digi
PREP HT
Voltage
230V ±5%
Power
3362W 40 and 20 posi-
tion
Power
3000W / 10 position
Internal thermal fuse
Manual reset above
450°C
Frequency
50/60 Hz
Ambient operating
temp.
Ambient to 450°C
RoHs
Toxic or hazardous substances and elements
Part name
Lead
(Pb)
Mer-
cury
(Hg)
Cadmium
(Cd)
Hexavalent
Chromium
(Cr
6+
)
Polybrominat-
ed Biphenyle
(PBB)
Polybrominated
Diphenyl Ether
(PBDE)
Housing
o
o
o
o
o
o
SiC Block
o
o
o
o
o
o
RTD probe
o
o
o
o
o
o
Over temperature
sensor
o
o
o
o
o
o
O
: Indicates that the toxic or hazardous substance
contained in all the homogenous materials for this
part, according to EIP-A, EIP-B, EIP-C, is below the
limit requirement in SJ/T11363-2006.
X
: Indicates that the toxic or hazardous substance
contained in all the homogenous materials for this
part, according to EIP-A, EIP-B, EIP-C, is above the
limit requirement in SJ/T11363-2006.
The main supply cord is constructed
of heat-resistant material and should
only be replaced by an equivalent or
better
cord.
Line voltage fl uctuations greater
than 10% will affect instrument
performance.
NOTE
NOTE
Summary of Contents for DigiPREP
Page 61: ...KeyPad Version 1 0 SCP SCIENCE 61 ELECTRICAL SCHEMATIC DigiPREP HT ...
Page 73: ...KeyPad Version 1 0 SCP SCIENCE 73 HEATER VERIFICATION ...
Page 83: ...KeyPad Version 1 0 SCP SCIENCE 83 ...
Page 84: ...SCP SCIENCE 84 KeyPad Version 1 0 ...
Page 85: ...KeyPad Version 1 0 SCP SCIENCE 85 ...