Figure C-1 Sprayer Nozzle Position
Item
Description
1
Corona discharge needle
2
Curtain plate
3
APCI probe
1. Use the previous horizontal and vertical micrometer settings or set them to 5 as a starting
position.
Note:
To avoid reducing the performance of the mass spectrometer, do not spray directly in
the aperture.
2. In the control software, monitor the signal or signal-to-noise of the analytes.
3. Use the horizontal micrometer to adjust the probe in small increments to achieve the best
signal or signal-to-noise ratio.
4. Use the vertical micrometer to adjust the probe in small increments to achieve the best
signal or signal-to-noise ratio.
5. Adjust the black electrode adjustment nut on the probe to move the electrode tube in or out
of the probe (to adjust the protrusion).
Ion Source Optimization
System User Guide
RUO-IDV-05-6475-C
SCIEX 3200 Systems
95/114