Clean the Curtain Plate
CAUTION: Potential System Damage. Do not rest the curtain plate or orifice plate on the
aperture tip. Make sure that the conical side of the curtain plate faces up.
CAUTION: Potential System Damage. To avoid damaging the aperture, do not insert a
wire or metal brush in the aperture on the curtain plate, orifice plate, or interface heater.
1. Pull the curtain plate off of the vacuum interface and then put it, conical side up, on a clean,
stable surface.
Figure 7-3 Curtain Plate Removal
The curtain plate is held in place by three retaining ball catches mounted on the orifice plate.
Tip!
If the curtain plate does not immediately separate from the orifice plate, then turn the
curtain plate slightly, less than 90 degrees, to release the ball spring latches.
2. Dampen a lint-free wipe with LC-MS-grade water and then clean both sides of the curtain
plate.
Note:
Use multiple wipes, as required.
3. Repeat step
using the cleaning solution.
4. Using a dampened wipe or small poly swab, clean the aperture.
5. Wait until the curtain plate is dry.
6. Inspect the curtain plate for solvent stains or lint, removing any residue with a clean, slightly
damp, lint-free wipe.
Service and Maintenance Information — Mass Spectrometer
SCIEX 3200 Systems