Prerequisites
• Make sure that all of the O-rings are present on the vacuum interface.
Figure 3-1 O-rings on the Vacuum Interface
Item
Description
1
Curtain plate
2
O-rings
1. Make sure that the ion source latches on either side of the ion source are pointing up in the
12 o’clock position. Refer to the section:
2. Align the ion source with the vacuum interface, making sure that the guide pins on the ion
source are aligned with the sockets in the vacuum interface.
3. Push the ion source gently against the vacuum interface and then rotate the ion source
latches down to lock the ion source in place.
The mass spectrometer recognizes the ion source and then shows the ion source
identification in the control software.
4. Connect the red PEEK tubing from the sample supply device to the other side of the
grounding union on the ion source.
Ion Source Installation
System User Guide
RUO-IDV-05-6475-C
SCIEX 3200 Systems
37/114