Table 9-1 System Issues (continued)
Symptom
Possible Cause
Corrective Action
The control software reports
that the mass spectrometer is
in Fault status because of the
ion source.
1. The probe is not installed.
2. The probe is not
connected securely.
Confirm the fault in the Status
panel of the device details
page.
1. Install the probe. Refer
2. Remove and then install
the probe. Tighten the
retaining ring securely.
Refer to the section:
.
The roughing pump is too hot
to touch.
The roughing pump filter trap
is blocked.
Contact the local QMP or
FSE.
The control software indicates
that the APCI probe is in use,
but the TurboIonSpray probe
is installed.
The F3 fuse is blown.
Contact an FSE.
The spray is not uniform.
The electrode is blocked.
Clean or replace the
electrode. Refer to the
document:
Turbo V Ion Source
Operator Guide
.
Clean or replace the
electrode. Refer to the
section:
The interface heater is not
ready.
The interface heater is faulty.
Contact the local QMP or
FSE.
The mass spectrometer
resolution is poor.
The mass spectrometer is not
tuned.
Use the Instrument
Optimization wizard to
optimize the mass
spectrometer. Refer to the
documents:
Software User
Guide
or
Help
.
Mass Spectrometer Troubleshooting
SCIEX 3200 Systems
78/114
System User Guide
RUO-IDV-05-6475-C