7. Troubleshooting
Problem: Cleaning solution leakage
Remedy
1. Ensure that the male outlet on the
cleaning solution bottle is tightly closed.
2. Ensure that the male and female
connectors have mated.
3. Check the cleaning solution tubing
for cracks and leaks. Replace if
necessary.
Problem: Water leaking from the door
(front of the unit)
Remedy:
1. Make sure unit is level.
2. Check adjustment on the door latch.
Loosen two screws and then slide door
latch in or out to adjust.
Problem: Instruments not dry
Remedy
:
1. Increase drying time
2. Replace Air filter
3. Replace dryer.
Dryer
Air
Filter
Male and
Female
Connectors
Cleaning
Solution
Tubing
Figure 30
Figure 32
Figure 31
Door Latch
Page 31
HYDR
IM
C51wd Service Guide