HYDR
IM
C51w Service Guide
11. Appendix
11.5 Changing the Air / Water Feed Lines
If for some reason there is an air or water
reduction to the LCS, the individual tubes
can be removed and replaced. To
remove the individual tubes, you will
need to first remove the door fascia.
(See Unit Overview for instructions on
removing the door fascia.)Once the door
fascia has been removed, remove the
individual tie-wrap connecting the tube
onto the LCS. Then follow the tube down
through the bottom pan and kickplate
until it reaches the manifolod (p/n 74-
111612)
If you need to remove the entire LCS
feed line assembly, then unscrew the
7 screws connecting the LCS to the
Hydrim door. Then follow the tubes down
through the kickplate and bottom pan to
the manifold. Disconnect the air check
valve and also the water filter check valve.
Then uncrew the two screws connecting
the manifold to the backet.
11.6 Removing the Solenoid
If the LCS solenoid needs to be replaced,
the top cover must first be removed.
(See Unit Overview for instructions on
removing the top cover.)
After the top cover has been removed,
disconnect the two tie-wraps connecting
the tubes to the solenoid.
Once the tubes have been disconnected
remove the two screws connecting the
solenoid to the bracket. Replace by
following the same instructions in
reverse.
solenoid
manifold
water
filter
check
valve
air
check
valve