Page 9
HYDR
IM
C51wd Service Guide
2.5 Drain Requirements
Connect the drain hose to the drain outlet. The drain hose can be attached to existing
drain lines using a 3.5 cm or larger standpipe / P-trap combination. If the hose is
connected directly to the drain line, fittings and adapters should not reduce water flow.
The drain hose should be attached to the main drain at a point no more than 1 metre.
above the base of the Hydrim. A floor drain is acceptable (check local codes).
2.6 Levelling the Hydrim
For the unit to function properly, it will need to
be correctly levelled. To level the unit, follow
these steps:
1.
Adjust the legs underneath the unit.
2.
Use the levelling bubble on the top right hand
side as a guide.
3.
When the bubble is in the centre, the unit is
correctly levelled.
2.7 Installing Cleaning Solution
Ensure that the quick-disconnect cap on the HIP Hydrim Cleaning Solution bottle is tight.
Install the bottle, and loosen the screw cap slightly (see Step 4) to prevent formation of a
vacuum in the bottle.
2. Installation
levelling bubble
Figure 5
Figure 6
screw cap
quick disconnect cap
MAX
MIN
Step 1
Step 2
Step 3
Step 4