![SciCan HYDRIM C15wd Service Manual Download Page 18](http://html1.mh-extra.com/html/scican/hydrim-c15wd/hydrim-c15wd_service-manual_1217208018.webp)
4.6 Removing The Door Fascia
To remove the door fascia, follow
these steps:
1.
Power OFF the unit and unplug.
2.
Open the door and remove
the 4 Phillips screws on the
perimeter of the door’s
inside face.
3.
Pull the door fascia towards
you and lift up.
4. Unit Overview
screws
Figure 19
Page 17
HYDR
IM
C51wd Service Guide
J6
J5
J2 J1
J3
J7
Figure 18f
Figure 18g
screws
screws
standoffs
printer
cable
6.
Disconnect the signal connector from the
bottom of the logic board (Figure 18e).
7.
Unclip the temperature sensor junction
block (not shown).
8.
Remove four screws (Figure 18f)
and lift I/O board off.
9.
Disconnect the printer cable from the
side of the logic board (Figure 18g).
10.
To remove logic board, remove upper
standoffs and two screws (Figure 18g).
11.
Reassemble in reverse order.