1. Introduction
1.1 Overview
This guide provides instructions for the servicing and repair of the Hydrim
®
C51wd
Instrument Washer-Disinfector. Every attempt has been made to provide accurate,
detailed instructions.
All servicing of the Hydrim C51wd should be done by certified personnel only. All local,
provincial, state and national regulations regarding the servicing of the class of device and
safety requirements must be observed.
Do not permit any person other than certified personnel to supply parts for, service, or
maintain a Hydrim C51wd. SciCan shall not be liable for incidental, special or
consequential damages caused by any maintenance or services performed on the
Hydrim C51wd by a third party, including lost profits, any commercial loss, economic loss,
or loss arising from personal injury.
The Hydrim C51wd Instrument Washer-Disinfector should only be installed and serviced
by a qualified contractor as it is an Installation Category 2 device. The contractor should
be experienced in installing equipment that requires electrical hook-up as well as
plumbing.
Hydrim C51wd Cycle Description Chart
** Cycle times depend on the temperature of incoming water.
Regular Cycle
Heavy Duty
Heavy Duty Cycle
Cycle
with Disinfection
Description
Use for moderately
Use for heavily
Use for heavily soiled
soiled loose
soiled instruments
instruments and
instruments
and cassette loads
cassette loads
Cold Prewash
<45ºC
<45ºC
<45ºC
Wash
50ºC
50ºC
50ºC
5 minutes
9 minutes
9 minutes
Rinse/Disinfect
60ºC
60ºC
80ºC for 10 minutes
Dry
0-15 min.
0-15 min.
0-15 min.
Total Time
**
19 minutes
23 minutes
45 minutes
without drying
Water
13 L
13L
13L
Consumption