4 Maintenance and servicing
16
SCHMIDT
®
ServoPress 605-660
– V1.1
Clean the ram with a lightly oiled cloth.
Clean the scale on the ram with a soft, lint-free cloth wetted with
isopropyl alcohol.
4.2 Adjusting the force measurement
Since the force-measuring system is usually used for measuring
characteristic quantities relevant to quality, it should be calibrated and, if
necessary, adjusted at reasonable intervals established by the owner. EN
ISO 9001 (Quality management systems - Requirements) makes a
calibration of the system mandatory at regular intervals.
Calibration is also offered as a service by the SCHMIDT Customer Service.
4.3 Lubricating the spindle
The lubrication of the spindle is stroke-dependent. Depending on the
configured operating profile, the point in time for the next lubrication is
calculated. When the given time has come, the control will request you to
lubricate.
For further information, refer to the "Instructions for Use - SCHMIDT Manual
workstation"
respectively
the
"Assembly
Instructions
-
SCHMIDT
Components for automation".
4.4 Filling the grease reservoir
WARNING
Shearing of finger(s)
The upward or downward movement of the used grease depot or spindle
nut creates a danger zone during operation of the press. Fingers can be
sheared off when reaching into this danger zone.
Mount the inspection cover on the inspection opening before switching
on the press system.
NEVER
operate the press without the inspection cover installed.
If the grease reservoir is almost empty, a message to refill the grease
reservoir is displayed.
To fill the grease reservoir, the ram of the press moves to a special position.
This happens automatically after answering the message.
If you are prompted ...
Switch off the press system with the main switch on the switch cabinet.
ServoPress 605-620
For the ServoPresses 605-620, the grease reservoir must be filled manually
using the lubrication nipple.
There is an inspection opening at the front of each press. Underneath is
Switching off the press
system