non-medical grade equipment and Listed / Approved / Certified Information Technology
Equipment (ITE) computer equipment is acceptable.
Protecting iPan Equipment from RF Interference
Although the iPan equipment is designed to provide a reasonable degree of protection
from electromagnetic interference, according to IEC International regulations, it must be
installed at an adequate distance from electricity transformer rooms, static continuity
units, two-way amateur radios and cellular phones. To ensure proper operation, the latter
can be used only at a minimum distance of 5 feet (1.5m) from any part of the equipment.
Any instrumentation or equipment for professional use located near iPan must conform to
Electromagnetic Compatibility regulations. Non-conforming equipment, with known
poor immunity to electromagnetic fields, may not operate properly unless they are
installed at a distance of at least 10 feet (3m) and supplied by a dedicated electric line.
Preventive Maintenance
There are no customer-serviceable components in the iPan system. However, before
operating the system, customers shall check it for any signs of physical damage or defect.
If detected, contact your local distributor of Schick Technologies products for further
instructions.
iPan User Guide
B1051601 Rev. –
iii