Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF
communication equipment should
be separated from the iPan system
by no less than the distances
calculated / listed below:
Recommended separation
distance:
d= (3.5/3)(Sqrt P)
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d= (3.5/3) (Sqrt P)
80 MHz to 800Mhz
d= (7/3) (Sqrt P)
800 MHz to 2.5Ghz
Where P is the maximum power
in watts (W) and D is the
recommended separation distance
in meters (m).
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey
should be less than the
compliance level (V1 and E1).
Interference may occur in the
vicinity of equipment containing
a transmitter.
Table 7. Recommended Separation Distance between Portable and Mobile RF
Communications Equipment and the iPan System
PLEASE NOTE: The iPan system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the iPan system can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the system as
recommended below, according to the maximum output power of the communications equipment.
Max
Output
Power
(Watts)
Separation (m)
150 kHz to 800 MHz
d=(3.5/V1)(Sqrt P)
Separation (m)
150 kHz to 800 MHz
d=(3.5/E1)(Sqrt P)
Separation (m)
800 MHz to 2.5 GHz
d=(7/E1)(Sqrt P)
0.01
0.12
0.12
0.23
0.1 0.37
0.37
0.74
1 1.17
1.17
2.33
10 3.69
3.69
7.37
100 11.7
11.7
23.33
iPan User Guide
B1051601 Rev. –