A-5. EMC Tables
Table 5. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
PLEASE NOTE: The iPan system is intended for use in the electromagnetic environment specified
below. The user of the iPan system must ensure that it is used in such an environment.
Emissions Test
Compliance
Guidance
RF emissions
CISPR 11
Group 1
ClassB
The iPan system uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Table 6. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
PLEASE NOTE: The iPan system is intended for use in the electromagnetic environment specified
below. The user of the iPan system must ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Electrical fast
transient/burst (EFT)
IEC 610004-4
±
2 kV for power
supply lines
±
1 kV for
input/output lines
±
2 kV for power
supply lines
±
1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2kV common
mode
±
1 kV differential
mode
±
2kV common
mode
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
(>95% dip )
for 0.5 cycle
(>60% dip)
for 5 cycles
(>30% dip)
for 25 cycles
(>95% dip)
for 5 sec
(>95% dip )
for 0.5 cycle
(>60% dip)
for 5 cycles
(>30% dip)
for 25 cycles
(>95% dip)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the iPan system requires
continued operation during mains
interruptions, it is recommended
that the iPan system be powered
from an uninterruptible power
supply or battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3A/m 3A/m
Power
frequency
megnetic
fields
should be at levels characteristic
of a typical location of a typical
commercial or hospital
environment.
B1051601 Rev. –
iPan User Guide
22