
Safety Issues
Equipment to be Operated and Serviced by Qualified Personnel Only
X-ray equipment produces ionizing radiation that may be harmful if not properly
regulated. It is therefore recommended that the equipment be operated by trained and
qualified personnel only, in accordance with all applicable local and federal regulations.
Only trained and qualified technicians are authorized to service this equipment. Power
supply lines must comply with safety legislation and have ground terminals for protective
earth connection. Always switch the equipment off and, if possible, disconnect it from
main power supply before cleaning the system.
Avoid Operating Equipment in the Presence of Interfering RF Devices and
Equipment
We recommend that you do not use the equipment in the presence of external
electromagnetic fields, even if compliant with specifications for electromagnetic
compatibility. Noise, generated by cellular phones, might interfere with the electronic
circuits of the system.
Apply Recommended Procedures for Cleaning the Equipment
Safe and proper operation of the equipment requires that a regular schedule of preventive
maintenance be followed. Refer to the Cleaning and Maintenance sections of this manual
for details.
Do Not Connect Items that are Not Part of the System
Only items specified for use with the equipment are to be connected to the system. The
equipment should not be used adjacent to other equipment that is not part of the system.
If, however, use with adjacent equipment is necessary, normal operation should be
observed and verified in that configuration.
Installers to Ensure that the iPan System Operates Optimally
Installers must ensure that the iPan system, when installed, provides the user with the
optimal use of the equipment. To verify this requirement, installers shall confirm that the
iPan system is installed as described in the Installation / Service Guide and shall perform
the appropriate procedures therein.
Take Appropriate Precautions during iPan Operation
Appropriate accessories, such as lead aprons, must be used, where necessary, to protect
the patient and the operator from radiation.
The iPan system has been determined to be in accordance with international safety
standards and is deemed suitable for use within the patient area, which extends from the
patient for a distance of 5 feet (1.5m). Outside the patient area, the presence of approved
B1051601 Rev. –
iPan User Guide
ii