5.6 Panoramic Calibration Properties
Please Note: Casual changes to the Acquisition Modules
section can adversely affect system performance. We do
not recommend making changes in this section unless
directed to do so by authorized support personnel.
•
Calibration File Path
(Displays the path where calibration files for
panoramic sensors are stored.)
•
Gap Width
(Displays the number of blank rows between the panoramic
sub-sensors.)
•
Calibration Threshold
(Displays the percentage above which raw pixels
are not calibrated.)
•
Dark Threshold
(Displays the threshold percentage above which dark
subtraction is not applied to raw pixels.)
•
Extra Bad Rows
(When checked, indicates that two additional blank rows
are marked bad and are added between the panoramic sub-sensors.)
•
Clip Saturated Pixels
(When checked, dark subtraction is modified for
pixels with values between 99 and 100%.)
•
Sirona Orthophos System
(When checked, accomodates the particular
design of this system in which the Sensor is inverted when installed.)