Overview
Specifications
Specifications
Built-in motorized
All models with the designation suffix ED...-CW, GK...,
calibration weight
GW.... or with a readability of 0.0001 g and all verified models
AC power source/power
AC adapter 230 V or 115 25 V,
requirements, voltage, frequency
+15% to – 20%, 48–60 Hz
Power consumption
VA
maximum 16; typical 8 (STNG6)
Approx. hours of operation with
the YRB05Z rechargeable battery
pack (backlighting on)
h
35
Ambient Conditions
The specifications given here are ensured under the following ambient conditions:
Operating temperature range
+10 to +30°C (273 to 303 K, 50 to 86°F)
Allowable ambient
operating temperature
+5 to +40°C (41 to 104°F)
Proper functioning is ensured within an ambient operating temperature range of 5 to 40°C (41 to 104°F).
Model-specific Specifications
Model
ED224S
ED124S
GK1203
GK703/
GK303
GK703-ST
Weighing capacity
220 g
120 g
1200 ct
700 ct
300 ct
Readability
0.0001 g0.0001 g0.001 ct
0.001 ct
0.001 ct
Tare rang
e (subtractive)
220 g120 g1200 ct
700 ct
300 ct
Repeatability (std. deviation)
<
±
0.0001 g0.0001 g0.001 ct
0.001 ct
0.001 ct
Linearity
<
±
0.0002 g0.0002 g0.002 ct
0.002 ct
0.002 ct
Response time (average)
s
2.5
2.5
1.5
1.5
1.5
Sensitivity drift within
<
±/K
2 · 10
–6
2 · 10
–6
2 · 10
–6
2 · 10
–6
2 · 10
–6
+10 to +30°C
Adaptation to ambient
By selection of 1 of 4 optimized filter levels;
conditions
display update: 0.1–0.4 (depends on filter level selected)
External calibration weight
(of at least accuracy class...)
g200 (E2)
100 (E2)
200 (E2)
100 (F2)
50 (F2)
Net weight, approx.:
kg
4.8
4.8
4.7
4.7
4.7
Weighing pan size
mm
90
d
90
d
90
d
90
d
/
35
d
90
d
Whg. chamber height
mm
230
230
160
160/38
160
Dimensions (WxDxH)
mm
230
+
303
+
330
230
+
303
+
260
Model GK703-ST:
230
+
303
+
138
54
Summary of Contents for ED124S