Recycling
Safety Inspection
If there is any indication that safe
operation of the balance/scale is no
longer warranted:
§
Turn off the power and disconnect the
equipment from AC power immediately.
> Lock the equipment in a secure place
to ensure that it cannot be used for the
time being.
Notify your nearest Sartorius Service
Center. Repair work must be performed
by trained service technicians.
We recommend having the power
supply inspected by a certified electrician
at regular intervals, according to the
following checklist:
– Insulating resistance: > 7 megaohms
measured with a constant voltage of at
least 500 volts at a 500 K-ohm load
– Leakage current: < 0.05 mA measured
with a properly calibrated multimeter
If you no longer need the packaging
after successful installation of the
equipment, you should return it for
recycling. The packaging is made from
environmentally friendly materials and
is a valuable source of secondary raw
material. Batteries are hazardous waste
and must be disposed of seperately.
Please deposit empty batteries in the
collection boxes set up in your area for
this purpose. On request, Sartorius can
provide GRS boxes for collecting used
batteries. (GRS stands for “Gemein-
sames Rücknahme System,” a German
organization for battery disposal.*)
Contact your local waste
disposal authorities if you
wish to scrap the equipment.
Sartorius AG in Goettingen
will take back equipment
and packaging for disposal in
accordance with the applicable laws.*
If you set up the equipment in a country
other than Germany, please contact
your local waste disposal authorities for
information on similar services.
* This service is offered only within
Germany