Example:
Calculating the weight per unit area of paper: An A4 sheet of paper is used in this example,
with surface dimensions of 0.210 m
+
0.297 m = 0.06237 m
2
. To determine the weight per
unit area, the total weight is divided by the surface.
Parameter settings:
Applic.
:
Calc..
:
Method
:
Div.
(menu code 2. 8. 1. 2)
Step
Key (or instruction)
Display/Data output
1. Tare the balance/scale
w
0.00
g
2. Activate divisor input
b
______0.
3. Set the divisor (in this example:0.06237):
Position the decimal point,
u
, 5x
b
,
__.00000
2
+
u
,
Enter numerals
Repeatedly
__.06000
or press and hold;
b
,
u
, etc.
__.06237
4. Store the divisor and
initialize the balance/scale
u
+
0.0
o
The current divisor remains
Div 0.6237
stored in battery-backed memory
until the setting is changed
5. Weight per unit area:
+
79.7
o
*
Place an A4 sheet of paper
on the balance/scale
6. If desired, print result
r
Res + 79.7 o
7. Toggle display between
Repeatedly:
b
+
4.97
g
*
weight and calculated value
+
79.7
o
*
8. Unload the balance/scale
+
0.0
o
*
9. Repeat as needed, starting from Step 5
31
Summary of Contents for ED124S
Page 3: ...Fran ais page xx Espa ol page xx Italiano page xx 2...
Page 60: ...59 C...