Example:
Counting parts of equal weight
Parameter settings:
Applic.
:
Count.
(menu code 2. 3.)
Step
Key (or instruction)
Display/Data output
1.
Place empty container
+
22.6
g
on the balance/scale
2.
Tare the balance/scale
w
0.0
g
3.
Add reference sample quantity
to container (in this example: 20 pcs)
4. Changing the reference
sample quantity:
b
ref
10
pcs
5. Select reference sample quantity:
Repeatedly:
b
In increments of 1 (1, 2, 3, etc. to 100) Press briefly
In increments of 10 (10, 20, etc. to 100)
b
press
ref
20
pcs
and hold
6.
Confirm selected reference sample
u
+
20
pcs
*
quantity and start application
nRef 20 pcs
The current reference weight remains
wRef 1.07 g
stored until a new reference is set
or the power supply is interrupted
7.
Add desired number of pieces
+
500
pcs
8.
If desired, print quantity
r
Qnt + 500 pcs
9.
Toggle display between
1.07
g
G
*
mean piece weight, weight,
+
535.0
g
*
quantity
Repeatedly:
b
+
500
pcs
*
10. Unload the balance/scale
–
21
pcs
*
11. Repeat as needed, starting from
Step 7
12. Delete reference value
c
0.0
g
27
Summary of Contents for ED124S