Troubleshooting Guide
Error codes are shown on the main display for approx. 2 seconds. The program then returns automati-
cally to the previous mode.
Display
Cause
Solution
No segments appear
No AC power is available
Check the AC power supply
on the display
The power supply is not
Plug in the power supply
plugged in
high
The load exceeds the balance/
Unload the balance/scale
scale capacity
low
or
Err 54
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
App.err.
Cannot store data:
Increase load
Load on weighing pan too
light or no sample on pan
while application is active
dis.err.
Data output not compatible
Change the configuration in
with output format
the operating menu
prt.err.
Interface port for printer
Reset the menu factory settings,
output is blocked
or
Contact your local Sartorius
Service Center
err 02
Calibration parameter
Calibrate only when zero
not met; e.g.:
is displayed
– balance/scale not tared
– Press
w
to tare the balance/scale
– load on weighing pan
– Unload the balance/scale
err 10
The
w
key is blocked when
Press
c
to clear the tare
there is data in the second tare
memory and release the tare key
memory (net-total); only 1 tare
function can be used at a time
err 11
Tare memory not allowed
Press
w
The weight readout
Unstable ambient conditions
Set up the balance/scale
changes constantly
(excessive vibration or draft)
in another area
at the place of installation
A foreign object is caught
Remove the foreign object
between weighing pan and
balance/scale housing
The weight readout
The balance/scale was not
Calibrate/adjust the balance/scale
is obviously wrongcalibrated/adjusted
Balance/scale not tared
Tare or zero the balance/scale
before weighing
before weighing
If any other errors occur, contact your local Sartorius Service Center.
Contact information:
Please point your Internet browser to: http://www.sartorius.com
51
Summary of Contents for ED124S
Page 3: ...Fran ais page xx Espa ol page xx Italiano page xx 2...
Page 60: ...59 C...