Example:
Determining residual weight in percent
Parameter settings:
Applic.
:
Percent
(menu code 2. 4.)
Reference percentage:
Ref 100
%
Step
Key (or instruction)
Display/Data output
1.
Tare the balance/scale
w
0.0
g
2. Information:
Enter reference percentage
b
ref 100
%
(Changing the reference:
see the previous page)
3.
Place sample equal to 100%
on the balance/scale
(in this example: 111.6 g)
4.
Initialize the balance/scale
u
+
100.0
%
*
The current reference weight
pRef 100 %
remains stored until a new reference
Wxx% 111.6 g
is set or the power supply
is interrupted
5.
Remove sample
(e.g., for drying)
6.
Place unknown weight on
+
94.9
%
*
balance/scale
(in this example: 322.5 g)
7.
If desired, print percentage
r
Prc + 94.9 %
8.
Toggle display between
+
105.9
g
*
weight and percentage
Repeatedly:
b
+
94.9
%
*
9.
Clear display of residual weight
and reference percentage
c
+
105.9
g
10. If desired, print net residual weight
r
N + 105.9 g
29
Summary of Contents for ED124S
Page 3: ...Fran ais page xx Espa ol page xx Italiano page xx 2...
Page 60: ...59 C...