Application Programs
Counting
Display symbol:
Z
Purpose
With the Counting program you can
determine the number of parts that each
have approximately equal weight. To do
this, a known number of parts (the refer-
ence sample quantity) is weighed first,
and the individual piece weight (reference
weight) is calculated from this result.
Thus the number of parts subsequently
placed on the balance/scale can be
determined from their weight.
Changing the Reference Sample Quantity
Activate function:
Press the
b
key
Select the desired reference sample quantity
(1 to 100):
In increments of 1: Press the
b
key
briefly
In increments of 10:
Press and hold the
b
key.
The quantity is stored in battery-backed
memory.
Reference Sample Updating
Automatic reference sample updating
optimizes the counting accuracy. You can
activate or deactivate this function in the
menu.
Automatic reference sample updating
is performed when the requirements,
including the specified stability criterion,
have been met.
The abbreviation
opt
, for “optimizing”,
is displayed briefly with the new reference
sample quantity.
Preparation
§
Select the Counting application in the
menu:
see “Configuration.”
§
Set the following parameters:
Applic.
Application program
Count.
Resolut.
Resolution
o
Disp.Acc.
Display accuracy
10-fold
10-fold higher
Ref.Updt.
Autom. ref. sample
updating
o
Off
Display accuracy
Autom.
Automatic
o = Factory setting
Printout: Counting
nRef
+
10
: Reference sample
quantity
wRef
+ 21.14 g
: Reference weight
Qnt
+
500 pcs
: Calculated quantity
26