Time Constant of Low Pass Filter
300ms
Roll Off
12 dB/oct
Sampling Rate
125Hz
Buffer 1
X
Buffer 2
Y
The step to finish the configuration in table 14 are as follows:
1. According to Table 14, choose the following options in
“
Input Setting
”
, including
“
Input Source
”
,
“
Input
Coupling
”
and
“
Input Notch Filter
”
with other options in default, as is shown in Fig.71. At last, click the button
to finish the configurations.
Fig.71 Input Configuration
2. Choose these options in reference signal configurations, including reference signal source type, external
reference and reference phase with other options in default, as is shown in Fig.72. At last, click the button
to finish the configurations.
If you don
’
t use
“
Auto Phase
”
and
“
Auto Scale
”
to set the phase shift and display scale automatically, do not click
the corresponding button
and just keep it unchangeable.
If you use
“
Auto Phase
”
to set phase shift angle automatically, you need to click the corresponding button
. When this button is clicked and it displays “being progressing”. The SE2031 will execute the auto
phase function. After finishing this function, this button will reset to the original status
. Now the
option
“
Phase
”
will update to the phase shift angle value that returned by SE2031.