<Voltage>
Setting the amplitude of sine output
In <Fixed> mode, users can set the amplitude of sine out through keyboard. The range is from 0.001 Vpp to 10
Vpp and the minimum resolution is 0.001Vpp.
<Sweep Set>
Setting frequency sweep parameter.
<Start>
the start frequency of sweep
<Stop>
the cutoff afrequency of sweep
<Step>
in <Linear> mode, the <Step> is the step frequency of sweep; in the <Log> mode, the <Step> is the
100 percentage frequency.
<Time>
the step time interval of sweep.
Combine the keyboard and the soft keys to set the start frequency, cutoff frequency, step Frequency and time. The
range of amplitude is from 0.001 Vpp to 5 Vpp. <Start> amplitude is 1.0 Vpp. in default and <Stop> amplitude is
5.000 Vpp in default. The minimum resolution of the amplitude is 0.001 Vpp. In the <Log> mode, the minimum
resolution of the step is 0.001% of the frequency; and the minimum/maximum resolution of the step time is
1ms/100,000ms.
<Sweep Run>
Setting amplitude sweep run mode.
<Stop>
stop sweep
<Single>
single sweep
<Loop>
loop sweep
When start from <Single> sweep, SE2031 does an internal amplitude sweep according to the settings of users.
After that, the status jumps back to <Stop> automatically.
When start from <Loop > sweep, SE2031 does constant loop sweep until users switch the status to <Stop>
manually.
3.3 [GAIN/TC]
You can choose [GAIN/TC] submenu in menus bar to enter this interface, as is shown in Fig.24.