7: 10 ms
16: 300 s
8: 30 ms
17: 1000 s
9: 100 ms
18: 3000 s
OFSL(?){i}
The OFSL command sets or queries the low pass filter slope. The parameter i selects
<6dB/oct> (i=0), <12dB/oct> (i=1), <18dB/oct> (i=2) or <24dB/oct> (i=3).
SYNC(?){i}
The SYNC command sets or queries the status of synchronous filter. The parameter i
selects Off (i=0) or On (i=1). The synchronous filtering is turned on only when the
reference frequency is less than 20 Hz.
4.2.5 Display and Output Commands
FPOP(?)i{,j}
The FPOP command sets and queries the output channel on the front panel of SE2031. The
parameter i must be set when sending this command. The parameter i selects CH1 (i=1) or
CH2 (i=2) as the output channel. The FPOP i, j command sets output i to
quantity j. More details are as follow.
CH1 (i=1)
CH2 (i=2)
j Input Type
j Output Type
0
R
0
R
1
X
1
X
2
Y
2
Y
3
θ
3
θ
4
Rh1
4
Rh1
5
Xh1
5
Xh1
6
Yh1
6
Yh1
7
θh1
7
θh1
8
Rh2
8
Rh2
9
Xh2
9
Xh2
10 Yh2
10 Yh2
11 θh2
11 θh2
12 Noise
12 Noise
13 E1
13 E1
14 E2
14 E2
15 E3
15 E3
16 E4
16 E4
OEXP(?)i{,x,j}
The OEXP i command sets or queries the output offsets and expands. The parameter i must
be set when sending this command. The parameter i selects the type of parameters below.