Fig.59 The Configuration Area of Input Signal
Table 3. Input Configuration
Input Source
Single-Ended Voltage
Differential Voltage
Current
Current Impedance 1MOhm Current Gain (when the input signal is small, the gain is 10
6
V/A )
100MOhm Current Gain (when the input signal is small, the gain is 10
8
V/A)
Input Shield
Grounding
Float (the input connector ground and the instrument ground are isolated
by a 10kΩ resistor.)
Ground (t the input connector ground and the instrument ground are
isolated by a 10 Ω resistor.)
Input Coupling
AC coupling
DC coupling
Input Notch Filter
None (Don’t open the notch filter)
Line Notch (start 50 Hz notch filter and suppress the frequency
interference)
Both Notch Filters (start 100 Hz Notch Filters and suppress the
frequency interference).
2x Line Notch (start 50 Hz and 100 Hz notch filters and suppress the
frequency and frequency second harmonic interference).
Note that:
After finishing all the settings, click
to finish the configuration for input signal. All
settings can be changed at the same time and you just need to click this button once.