![Saluki SE2031 User Manual Download Page 79](http://html1.mh-extra.com/html/saluki/se2031/se2031_user-manual_3700286079.webp)
Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
79
7. Filter Configuration
The filter configuration is in the red box in the Fig.64. The details of configuration are shown in Table 8.
Fig.64 The Configuration Area of Filter Configuration
Table 8. Filter Configuration
Time Constant
10μs
3ms
1s
300s
30μs
10ms
3s
1000s
100μs
30ms
10s
3000s
300μs
100ms
30s
1ms
300ms
100s
Filter Slope
6dB/oct
12dB/oct
18dB/oct
24dB/oct
SYNC Filter
Disable
Enable
Note that:
After choosing all the settings, click the button
to finish the configuration; all settings can be
changed at the same time and you just need to click this button once.
After changing the settings, you need to click this button to finish this configuration. Otherwise the settings are
void.
The configurations of
“
Sync Filter
”
is relatively independent. Click the corresponding
not the
button
to close or open the auto function. The button will switch to
when
click the button
.This means the SE2031 is executing the auto function. If you need to close the