Figure 4-4 Calibration Switch is on
4.2.
Embedded calibration kit
4.2.1.
Introduction
Operators can quickly perform “one-click” single-port calibration by embedded electronic calibration, which is not only of simple operation
and high precision calibration, but also able to adapt to the environment temperature of 0
℃
to +50
℃
, significantly improving the test
efficiency. Meanwhile, the configuration of embedded electronic calibration of the test does not affect the use and performance of
mechanical calibration kits. They can be used to meets the testing requirements of a variety of occasions.
4.2.2.
Calibration Procedure
The embedded calibration process is as follows:
1) Before performing embedded calibration, please keep the measurement port is not connected.
2) When parameter settings are done, press
【
Cal
】
key, click [Embedded Calibration] menu item as shown in Figure 4-5, the
instrument will automatically perform a calibration procedure with a prompt saying “Embedded Calibration Measuring……”
displayed in the interface.