Operating concepts
R&S
®
NRP18S-xx
29
User Manual 1178.3686.02 ─ 07
5 = R&S
NRP
‑
ZK8 cable
6 = Sensor input connector of the R&S
NRX
7 = R&S
NRX base unit
1.
NOTICE!
Incorrectly connecting or disconnecting the power sensor can damage
the power sensor or lead to erroneous results. Ensure that you connect or discon-
nect the power sensor as described in
Chapter 3.4, "Connecting to a DUT"
on page 13.
Connect the power sensor to the signal source.
2. Connect the cables as shown in
Starting a measurement
For a detailed description of how to measure in this setup, refer to the user manual of
the R&S
NRX.
1. Preset the R&S
NRX and the connected R&S
power sensors.
a) Press the [Preset] key.
b) Tap "Preset".
All parameters are set to their defaults.
2. If you use the power sensor without the attenuator supplied in the delivery or with
another 2-port device, adjust the S-parameter correction. See
3.
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [Zero] key of the R&S
NRX.
c) Tap "Zero All Sensors".
4. Configure the measurement.
a) In the "Measurement Settings" dialog, select the "Measurement Type", for
example "Continuous Average".
b) Tap "Quick Setup"
> "Auto Set".
5. Switch on the signal source.
The measurement starts, and the result is displayed in dBm.
6. If necessary, perform further settings.
R&S
NRX