Remote control commands
R&S
®
NRP18S-xx
105
User Manual 1178.3686.02 ─ 07
Static errors occur when you select conflicting settings. For example in timeslot mode,
a static error occurs with the following settings:
●
Width of a timeslot: 100
µs
●
Exclude time at the start of the slot: 40
µs
●
Exclude time at the end of the slot: 60
µs
Then there is "nothing left" to be measured, and a static error appears.
Static errors, as a rule, prevent the execution of normal measurements.
Errors in SCPI communication are queried using
.
Usage:
Query only
SYSTem:SERRor:LIST:ALL?
Queries all changes in the static error queue that have not been read yet and removes
them from the queue.
Example:
SYST:SERR:LIST:ALL?
Response:
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification",0,"removed at uptime:2944;
notice; auto-averaging exceeded maximum time;
Notification"
.
Usage:
Query only
SYSTem:SERRor:LIST[:NEXT]?
Queries the list of static error changes for the oldest entry and removes it from the
queue.
Returns an error number and a short description of the error.
Example:
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Usage:
Query only
SYSTem:SUTime
<update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode,
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results.
Configuring the system