Operating concepts
R&S
®
NRP18S-xx
27
User Manual 1178.3686.02 ─ 07
Setup
1
2
3
4
5
6
7
Figure 5-2: Setup with the R&S
Power Viewer
1 = Signal source
2 = Attenuator
3 = R&S NRP18S-xx power sensor
4 = Host interface connector
5 = R&S
NRP
‑
ZKU cable
6 = USB connector
7 = Computer with installed R&S
Power Viewer
1.
NOTICE!
Incorrectly connecting or disconnecting the power sensor can damage
the power sensor or lead to erroneous results. Ensure that you connect or discon-
nect the power sensor as described in
Chapter 3.4, "Connecting to a DUT"
on page 13.
Connect the power sensor to the signal source.
2. Connect the cables as shown in
For a detailed description, refer to
Chapter 3.7.1.1, "Simple USB connection"
Starting a measurement
For a detailed description, refer to the operating manual of the R&S
Power Viewer. The
manual is installed automatically during the installation of the R&S
Power Viewer.
1. Start the R&S
Power Viewer.
2. If you use the power sensor without the attenuator supplied in the delivery or with
another 2-port device, adjust the S-parameter correction. See
3. Execute zeroing.
Note:
Turn off all measurement power signals before zeroing. An active measure-
ment signal during zeroing causes an error.
4. Switch on the test signal of the signal source.
5. Select a measurement.
6. Start the measurement.
R&S
Power Viewer