Remote control commands
R&S
®
NRP18S-xx
81
User Manual 1178.3686.02 ─ 07
3. Make sure that the S
-
parameter settings
—
selected S
-
parameter device, S
-
param-
eter correction state
—
always match the used hardware configuration.
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 8-7: Operation with 2
-
port device between signal source and power sensor input
Configuring the S
-
parameter correction
[SENSe<Sensor>:]CORRection:SPDevice:LIST?
................................................................ 81
[SENSe<Sensor>:]CORRection:SPDevice:SELect
.............................................................. 81
[SENSe<Sensor>:]CORRection:SPDevice:STATe
............................................................... 81
[SENSe<Sensor>:]CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been downloaded to the power
sensor. The result of the query indicates the consecutive number and mnemonic of
each data set.
Usage:
Query only
[SENSe<Sensor>:]CORRection:SPDevice:SELect
<num>
Selects a downloaded data set for S-parameter correction.
See also
Parameters:
<num>
Range:
1 to 1999
*RST:
1; can differ if a calibration set defines another
value.
[SENSe<Sensor>:]CORRection:SPDevice:STATe
<state>
Enables or disables the S-parameter correction. If activated, uses the S-parameter
data set selected by
[SENSe<Sensor>:]CORRection:SPDevice:SELect
See also
Parameters:
<state>
*RST:
OFF; can differ if a calibration set defines another
value.
Example:
CORR:SPD:SEL 1
Selects an S-parameter correction data set.
CORR:SPD:STAT ON
Enables the S-parameter correction.
Configuring basic measurement parameters