![Ronde & Schwarz 1424.6721K02 User Manual Download Page 14](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980014.webp)
Preparing for use
R&S
®
NRP18S-xx
14
User Manual 1178.3686.02 ─ 07
To connect to the DUT
1. Ensure that the RF connector of your DUT is compatible with the RF connector of
the power sensor or attenuator.
See
.
2. Inspect both RF connectors carefully. Look for metal particles, contaminants and
defects.
If either RF connector is damaged, do not proceed, because the risk of damaging
the mating connector is too high.
See also
Chapter 13.1, "Regular checks"
3. Insert the RF connector straight into the RF output of your DUT. Take care not to tilt
it.
180
4.
NOTICE!
Risk of damaging the center pin of the RF connector. Only rotate the hex
nut of the RF connector. Never rotate the power sensor itself.
Tighten the RF connector manually.
To disconnect from the DUT
►
NOTICE!
Risk of damaging the center pin of the RF connector. Always rotate only
the union nut of the RF connector. Never rotate the power sensor itself.
If the attenuator is attached to the power sensor:
a) Carefully loosen the union nut of the RF connector at the input of the attenua-
tor.
b) Remove the sensor with the attenuator.
Normally, it is not necessary or advisable to separate the attenuator from the sen-
sor if both are operated together.
► If you use the sensor without attenuator:
a) Carefully loosen the union nut at the RF connector of the sensor.
b) Remove the sensor.
3.5
Powering the power sensor
The electrical power for the R&S NRP18S-xx is supplied over the host interface. See
Powering the power sensor