Operating concepts
R&S
®
NRP18S-xx
26
User Manual 1178.3686.02 ─ 07
Connect the power sensor to the signal source.
2. Connect the power sensor to the computer as shown in
.
For a detailed description, refer to
Chapter 3.7.1.1, "Simple USB connection"
Starting a measurement
For a detailed description of how to measure in this setup, refer to the operating man-
ual of the R&S
NRPV.
1. Start the R&S
NRPV.
2. If you use the power sensor without the attenuator supplied in the delivery or with
another 2-port device, adjust the S-parameter correction. See
3. Execute zeroing.
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
4. Switch on the test signal of the signal source.
5. Start a measurement.
5.4
R&S
Power Viewer
The R&S
Power Viewer is software that simplifies many measurement tasks. It is provi-
ded on your documentation CD-ROM and on the Rohde & Schwarz website as a sepa-
rate standalone installation package.
Required equipment
●
R&S NRP18S-xx power sensor
●
R&S
NRP
‑
ZKU cable or an R&S
NRP
‑
Z5 sensor hub and an R&S
NRP
‑
ZK6 cable
to connect the sensor to the computer
●
Computer with installed:
–
R&S
NRP Toolkit V
4.20 or higher
–
R&S
Power Viewer version 9.2 or higher (refer to the operating manual of the
R&S
Power Viewer for a description of the installation process)
If you want to use an android device like a tablet or a smartphone, use the R&S
Power
Viewer Mobile. For details, see
R&S
Power Viewer