Instrument Functions
R&S
®
NGL200/NGM200
49
User Manual 1178.8736.02 ─ 09
The R&S
NGL/NGM displays the "Output Impedance" dialog.
2. Set the required value.
The R&S
NGL/NGM displays the onscreen keypad for entry.
3. Confirm value with the unit keys.
The output impedance control loop depends on the setting "Fast Transient Response".
See
Chapter 6.2.2, "Fast Transient Response"
The output impedance is only active during CV (constant voltage) operation mode.
6.2.3.2
Delay
1. Select the "Delay" menu item to configure the required values.
The R&S
NGL/NGM displays the "Output Delay" dialog.
2. Set the required value.
The R&S
NGL/NGM displays the onscreen keypad for entry.
3. Confirm value with the unit keys.
The output delay is the time between the "Output On" event and the available voltage
at the output terminals. See
.
Activating the Channels Output