Remote Control Commands
R&S
®
NGL200/NGM200
172
User Manual 1178.8736.02 ─ 09
Usage:
Query only
HCOPy:SIZE:Y?
Returns the vertical dimension of the screenshots.
Usage:
Query only
LOG[:STATe]
<state>
LOG[:STATe]?
Sets or queries the data logging state.
Parameters:
<state>
1
Data logging function is enabled.
0
Data logging function is disabled.
*RST:
0
Example:
LOG ON
LOG?
-> 1
Data logging function is activated.
LOG:COUNt
<count>
LOG:COUNt?
[<Return min or max>]
Sets, queries the number of measurement values to be captured.
Setting parameters:
<count>
<numeric value> | MIN | MAX
<numeric value>
Number of measurement values to be captured is set in the
range of 1 to 10000000.
MIN
Minimum number of measurement values to be captured is set
at 1.
MAX
Maximum number of measurement values to be captured is set
at 10000000.
Parameters for setting and query:
<count>
MIN | MINimum | MAX | MAXimum
Returns the number of measurement values.
Example:
LOG:COUN MAX
LOG:COUN? MAX
-> 10000000
Data and File Management Commands