Remote Control Commands
R&S
®
NGL200/NGM200
105
User Manual 1178.8736.02 ─ 09
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe:CHANnel
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe:CHANnel?
<arg0>
Sets or queries the device channel for trigger source "Operation Modes".
See
Parameters for setting and query:
<arg0>
OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2 | ANY
OUT1 | OUTP1 | OUTPut1
Ch1 is selected as the device channel for trigger source.
OUT2 | OUTP2 | OUTPut2
Ch2 is selected as the device channel for trigger source.
ANY
Any of the available channel which meets the condition sets as
the trigger source.
Example:
TRIG:SOUR:OMOD:CHAN 1
Ch1 is selected as the device channel for trigger source "Opera-
tion Modes".
TRIGger[:SEQuence][:IMMediate]:SOURce:OUTPut:CHANnel
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:OUTPut:CHANnel?
<arg0>
Sets or queries the device channel for trigger source "Output".
See
Parameters for setting and query:
<arg0>
OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2 | ANY
OUT1 | OUTP1 | OUTPut1
Ch1 is selected as the device channel for trigger source.
OUT2 | OUTP2 | OUTPut2
Ch2 is selected as the device channel for trigger source.
ANY
Any of the available channel which meets the condition sets as
the trigger source.
Example:
TRIG:SOUR:OUTP:CHAN 1
Ch1 is selected as the device channel for trigger source "Out-
put".
7.5
Configuration Commands
The following subsystems contain the commands for channel selection, voltage and
current settings for the instrument.
Configuration Commands