Remote Control Commands
R&S
®
NGL200/NGM200
124
User Manual 1178.8736.02 ─ 09
OUTPut:IMPedance
<resistance>
OUTPut:IMPedance?
[<Channel list>]
Sets or queries source impedance for the signal specified in ohms.
Parameters:
<arg0>
<numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault | <list>
<numeric value>
Numeric value of the impedance ohm.
MIN | MINimum
Minimum value of the impedance at -0.05 ohms.
MAX | MAXimum
Maximum value of the impedance at 100 ohms.
DEF
Default value of the impedance at 0 ohms.
*RST:
0
Default unit: ohm
Parameters for setting and query:
<Channel list>
<list>
Example:
OUTPut:IMPedance 1
OUTPut:IMPedance?
-> 1
Returns output impedance of 1 ohm.
Example:
OUTPut:IMPedance? (@1)
Returns output impedance at channel 1.
OUTPut:IMPedance:STATe
<state>
OUTPut:IMPedance:STATe?
[<Channel list>]
Sets or queries the impedance target for the selected channel.
Parameters:
<state>
0
Deactivates output impedance for the selected channel.
1
Activates output impedance for the selected channel.
Parameters for setting and query:
<Channel list>
<list>
Example:
OUTPut:IMPedance 1
OUTPut:IMPedance:STAT?
-> 1
Returns output impedance state as on.
Example:
OUTPut:IMPedance:STAT? (@1)
Returns output impedance state at channel 1.
Configuration Commands